37
Setup
Calibrating the sensors
The sensors and analogue inputs can be calibrated using a 2 point calibration
process. Those sensors marked “not changeable” in the Calibration column
cannot be calibrated.
Calibration and adjustment of the HygroClip2
probe is carried out using the ROTRONIC
HW4 software.
Select User settings and at the bottom click on Read user settings. Under
General > Calibration, check the Change values checkbox.
Choose Channel
3
.
Select a channel and then click Calibration.
3
4
Calibration can be carried out a) manually or b) by measuring the signal and sub-
sequently entering the target value.
a) Manual calibration
First enter the actual value and then the associated target value for points P1 and
P2. Activate the completed calibration by checking the checkbox
4
. Confirm
your selection with OK.
b) Calibration using the measurement function
Apply the desired signal for point P1 and click Measure. The
measured actual value is displayed. Enter the associated tar-
get value for point P1. Repeat this procedure for P2. Activate
the completed calibration by checking the checkbox
4
. Con-
firm your selection with OK.
Activate the completed calibration by checking the checkbox
4
and then click
OK. If necessary, calibrate further sensors (channels). With Write user settings
2
you can simultaneously transfer all of the completed calibrations to the logger.